Description
KEELER APPLANATION TONOMETER D- KAT
Specification
Keeler digital T type for Haag Streit
type illumination systems Fitted to the guide plate on the optical axis for the microscope and illumination unit arm
Keeler digital R type for some Mountable on post on microscope
Zeiss type and Haag Streit type
illumination systems
Installation Assembled on the guide plate on the slit lamp arm or mounting post
Measurement range 5–65 mmHg (0.66 – 8.46kPa)
Measurement deviation 0.49 mN or 1.5% of measurement value, whichever is the greater
Operating temperature range From 10ºC to 35ºC
Reverse span (Hysteresis) ≤ 0.49 mN
Net weight
Keeler digital T type 0.37 kg (without accessories)
Keeler digital R type 0.58 kg (without accessories)
Part Numbers
Keeler digital T type 2414-P-2032
Keeler digital R type 2414-P-2042
Environmental
Ambient Conditions
Transportation Temperature -40ºC to +70ºC
Air Pressure 500 hPa to 1060 hPa
Relative humidity 10% to 95%
Storage Temperature -10ºC to +55ºC
Air Pressure 700 hPa to 1060 hPa
Relative humidity 10% to 95%
Use Temperature +10ºC to +35ºC
Air Pressure 800 hPa to 1060 hPa
Relative humidity 30% to 75%
Down load D-KAT Brochure
http://www.mediscience.co.in/wp-content/uploads/2018/02/D-KAT-A4-spec-sheet-reprint-Dec-13.pdf
Reviews